Meaning
Physical contact between a measurement probe and a part surface causes a microscopic indentation or bending that can introduce error into the final reading. Stylus force deformation occurs because the spring pressure required to trigger the sensor also exerts a load on the material being measured. This effect is particularly significant when inspecting soft materials like plastics or thin-walled components that lack structural stiffness.
Force Magnitude
Pressure applied by the ruby or diamond tip ranges from a few grams to several Newtons depending on the probe design. In stylus force deformation, the local stress at the point of contact can exceed the elastic limit of the material, leaving a permanent mark. Even if the material recovers, the momentary deflection means the recorded position is slightly deeper than the actual surface.
Minimizing this error requires the use of low-force sensors or larger diameter styli.
Material Influence
Stiffness of the part determines how much it will move under the weight of the measurement probe. When calculating stylus force deformation, engineers must consider the Young’s modulus of the part material. Aluminium resists the probe much better than a soft polymer or a thin sheet of glass.
If the deformation is predictable, it can be compensated for in the measurement software to improve the accuracy of the result.
Precision Impact
High accuracy metrology in the micron range is impossible without accounting for these contact forces. Stylus force deformation represents a hidden variable that can consume a significant portion of a tight tolerance. By selecting the correct probe force for the application, a laboratory can ensure that its measurements are both repeatable and accurate.
This level of care is required for the validation of precision medical and aerospace components.