Meaning
Calibration standards consisting of a series of uniform steps with increasing thicknesses provide a baseline for measuring x-ray absorption and grey-value variations. A step wedge phantom is scanned to correlate the reconstructed voxel intensity directly with the known material thickness. This correlation enables quantitative beam hardening correction and density calibration.
It is an essential tool for maintaining the traceability of industrial CT scanners.
Calibration Reference
Industrial scanners must be calibrated regularly to ensure that grey values remain consistent over time. Scanning a step wedge phantom produces a staircase-like intensity profile on the detector. This profile is used to calculate the relationship between material thickness and x-ray attenuation.
This calculation helps to correct for the non-linear absorption of polychromatic x-ray beams.
Greyscale Correlation
Reconstruction software uses the grey values of the scan to determine the local density of the sample. By comparing these values to the known steps of the step wedge phantom, the software can build a precise calibration curve. This calibration allows for the accurate measurement of porosity and material distribution.
It is particularly useful when scanning multi-material components where different phases must be separated. This step ensures that the reconstructed volume accurately represents the physical sample.
Density Validation
Verifying the density of a scanned object requires a known reference scanned under identical conditions. A step wedge phantom made of the same material as the test piece provides this reference. This setup allows the inspector to verify that the density is accurate.